SANIMED G
Description:
The vertical bedhead unit SANIMED G has a structure inside which there is a mechanical separation dividing the medical gases distribution from the electrical services and the other services channel. The medical gases outlets and the electrical sockets are located in one side in the version for one bed and in two sides in the version for two beds. This configuration can be customized according to the Customer specific needs.
General information
- Body: Supporting structure made of steel and frontal panel made of MDF
- Maximum height: 2000 mm
- Protection degree IP: IP20 general - IP40 lighting circuit - IP21 electrical sockets
- Standard colour: Opaque RAL9010
- SURFACE TREATMENT ON REQUEST: Anodization, antimicrobial coating, decorative finishing. Check the possible finishings
- Electrical equipment: White UNEL 2x10 / 16A + T current socket, 3-module decorative plate (UNEL + data socket), nos.2 4-module decorative plate with switches and data sockets, nos.2 equipotential sockets.
- MEDICAL GAS SYSTEM: 3 preparation (holes) for connection of the terminal units for the passage of medical gases (type according to the customer's technical specification)
- Feeding tension: 230V / 50Hz
- Additional feeding tension: 24V AC
- Maximum socket power: 3000W
For specific needs, please contact our Sales Department.
CLASSIFICATION ACCORDING TO THE 93/42/CEE DIRECTIVE
- I class (no gas) - IIa/IIb (with gas)
- Non-stop operation
- No sections are applie
- Suitable for the installation in group 0, 1 and 2 medical premises
- Not suitable for the installation in explosion risk environments (locations with gases and flammable vapours)
- Not suitable for the installation in environments with an oxygen amount superior to the 25% with an environmental pressure until the 110kPa in usual circumstances
- Not suitable for the categories AP and APG provisions
REFERENCE DIRECTIVES AND REGULATIONS
93/42/CEE; UNI EN ISO 11197; CEI EN 60601-1; CEI EN 60598; CEI EN 55015; CEI EN 61547; UNI CEI EN ISO 14971; UNI CEI EN ISO 13485
- I class (no gas) - IIa/IIb (with gas)
- Non-stop operation
- No sections are applie
- Suitable for the installation in group 0, 1 and 2 medical premises
- Not suitable for the installation in explosion risk environments (locations with gases and flammable vapours)
- Not suitable for the installation in environments with an oxygen amount superior to the 25% with an environmental pressure until the 110kPa in usual circumstances
- Not suitable for the categories AP and APG provisions
REFERENCE DIRECTIVES AND REGULATIONS
93/42/CEE; UNI EN ISO 11197; CEI EN 60601-1; CEI EN 60598; CEI EN 55015; CEI EN 61547; UNI CEI EN ISO 14971; UNI CEI EN ISO 13485
TELL YOUR NEEDS
We are specialists in designing customized lighting solutions in a short time.
If you are an architect, a designer or a manager, contact us to explain your needs in detail.
We are specialists in designing customized lighting solutions in a short time.
If you are an architect, a designer or a manager, contact us to explain your needs in detail.